Customization: | Available |
---|---|
Type: | IVD Instrument |
Throughput: | 96 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model | Throughput | Gradient | Channel 1 | Channel 2 | Channel 3 | Channel 4 | Channel 5 | Channel 6 |
FAM, SYBR Green I, SYTO 9, Eva Green, LC Green |
HEX, VIC, TET, JOE |
ROX, Texas Red |
Cy5 | Alexa Fluor 680 | FRET | |||
Gentier 96E | 1-96 | Yes | √ | √ | √ | √ | √ | √ |
Model | Gentier 96E |
Throughput | 1- 96 |
Fluorescence Channels | 6 |
Fluorescence Scanning Time | 7s |
Dye Compatibility | Channel 1: FAM, SYBR Green I , SYTO 9, Eva Green, LC Green Channel 2: HEX, VIC, TET, JOE Channel 3: ROX, Texas Red Channel 4: Cy5 Channel 5: Alexa Fluor 680 Channel 6: FRET |
Optical System | |
Light Source | High-brightness, long life and maintenance-free LED light source,excitation from the top |
Detector | Photodiodes (PDs), top scanning |
Excitation Range | CH1: 465nm CH2:527nm CH3:580nm CH4:632nm CH5:680nm CH6:465nm |
Detection Range | CH1: 510nm CH2:563nm CH3:616nm CH4:664nm CH5:730nm CH6:616nm |
Fluorescence Dynamic Range | Adjustable |
Sample Dynamic Range | From 101 to 1010copies |
Thermal Black | |
Heating Method | Peltier |
Heating Rate | ≥6.1 ºC/s |
Cooling Rate | ≥5.0 ºC/s |
Temperature Uniformity | ±0.1 ºC |
Temperature Accuracy | ≤0.1 ºC |
Gradient Range | 1ºC-40 ºC |
Gradient Block | 12 row |
Special Temperature Protocol | Thermal gradients PCR, long PCR, Touch Down PCR |
Sample Testing Linearity and Repeatability | Linear correlation: /r/ ≥ 0.999 Repeatability: cycle threshold (Ct) value CV≤0 0.5% |
Software Functions | |
Control Modes | Model 1: 10.4 inch touch screen Mode 2: PC direct control |
Power Failure Protection | Automatically start running experiments after power supply, no need to wait PC software |
Data Storage and Transmission | Upload and download through USB disk. 1000 results can be stored in machine |
Reporting Function | Templates reserved; customized experiment report |
Key Applications | Relative quantification, absolute quantification, melting curve analysis ,SNP analysis |
Others | |
Operating System for PC | Win7/Win10/Win11 |
Power Supply and Power Consumption | AC 100-240V,50-60HZ, 900VA |
Weight | 30 Kg (net) |
Instrument Dimension | 555*475*484mm (W*L*H) |
Suitable Consumables | 0.2mL 96-well plates, 8-tube strips, single tubes(clear, frosted and white). |